Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Published Jul 2, 2018

Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Affected products

Remediation

Vendor solution

Medtronic will release several rolling over-the-air product updates that will mitigate the vulnerabilities described within this advisory. These updates will be applied to devices automatically as part of standard, reoccurring update processes. In addition, Medtronic has increased security monitoring of affected devices and related infrastructure.

Medtronic has released additional patient focused information, at the following location:

https://www.medtronic.com/security