Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers Missing Encryption of Sensitive Data
Published Dec 14, 2018
4.6
MEDIUMCVSS 3.1
EPSS 0.32%
Description
Medtronic CareLink and Encore Programmers
do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .
Affected products
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Medtronic | 29901 Encore Programmer | unaffected |
| ||||||
| Medtronic | CareLink 2090 Programmer | unaffected |
| ||||||
| Medtronic | CareLink 9790 Programmer | unaffected |
|
Configuration 1
Running on/with
- n/a
Configuration 2
Running on/with
- n/a
Configuration 3
Running on/with
- n/a
No data.
No Red Hat product state for this CVE.
No package ranges for this CVE.
Remediation
Vendor solution
The CareLink 9790 Programmer has been placed into end-of-life status and is no longer supported by Medtronic. Medtronic recommends users no longer use the 9790 for any purpose.
The CareLink 2090 and 29901 Encore programmers store PHI and PII as part of their normal operating procedure. Medtronic recommends that when devices are storing PHI/PII it should be retained on these programmers for the least amount of time necessary, and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy. Please contact a Medtronic representative for proper disposal and PHI/PII retention setting instructions.
All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.
Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, hospitals and clinicians should:
* Maintain strict physical control of the programmer. * Use only legitimately obtained programmers and not ones provided by any third party.
Proper disposal of these programmers and the associated electronic media storing data is critical for the continued protection of any PHI and PII residing on the programmer.
Medtronic has released a security bulletin related to this advisory that is available, with contact information, at the following location:
https://www.medtronic.com/security
Metrics
No CVSS v4.0 score for this CVE.
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
No CVSS v3.0 score for this CVE.
AV:L/AC:L/Au:N/C:P/I:N/A:N
This CVE is not in the KEV list.
No CISA SSVC assessment for this CVE yet.
Estimated probability of exploitation in the wild in the next 30 days (FIRST EPSS). As of Oct 3, 2026.
Score over time
2021–2026- EPSS v1
- EPSS v5
- EPSS v2
- EPSS v3
Percentile over time
- EPSS v1
- EPSS v5
- EPSS v2
- EPSS v3
Table of values (9 key points)
Flat stretches are collapsed; showing up to 120 newest points.
| Date | Score | Percentile | Model |
|---|---|---|---|
| Oct 3, 2026 | 0.32% (0.00315) | 22.25th | v5 (v2026.06.15) |
| Sep 20, 2026 | 0.32% (0.00315) | 24.69th | v5 (v2026.06.15) |
| Jul 20, 2024 | 0.05% (0.00053) | 21.45th | v3 (v2023.03.01) |
| May 6, 2024 | 0.05% (0.00053) | 20.06th | v3 (v2023.03.01) |
| Mar 7, 2023 | 0.06% (0.00064) | 26.02th | v3 (v2023.03.01) |
| Mar 6, 2023 | 0.89% (0.00885) | 27.89th | v2 (v2022.01.01) |
| Feb 4, 2022 | 0.89% (0.00885) | 10.50th | v2 (v2022.01.01) |
| Feb 3, 2022 | 0.62% (0.00624) | 17.80th | v5 (v2026.06.15) |
| Apr 14, 2021 | 0.62% (0.00624) | 0.00th | v1 |
References (3)
- http://www.securityfocus.com/bid/106215 vdb-entryThird Party AdvisoryVDB Entry
- https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-9790-2090-29901.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-347-01 Third Party AdvisoryUS Government Resource
| Link | Providers | Tags |
|---|---|---|
| http://www.securityfocus.com/bid/106215 | vdb-entryThird Party AdvisoryVDB Entry | |
| https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-9790-2090-29901.html | ||
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-347-01 | Third Party AdvisoryUS Government Resource |
Change history (0)
No recorded changes yet.