Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Cleartext Transmission of Sensitive Information
Published Aug 13, 2018
4.8
MEDIUMCVSS 3.1
EPSS 0.47%
Description
Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.
Affected products
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Medtronic | MMT – 511 pump Paradigm | unaffected |
| ||||||
| Medtronic | MMT – 512 / MMT – 712 Paradigm x12 | unaffected |
| ||||||
| Medtronic | MMT – 515 / MMT – 715 Paradigm x15 | unaffected |
| ||||||
| Medtronic | MMT – 522 / MMT – 722 Paradigm REAL-TIME | unaffected |
| ||||||
| Medtronic | MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME | unaffected |
| ||||||
| Medtronic | MMT – 523 / MMT – 723 Paradigm Revel | unaffected |
| ||||||
| Medtronic | MMT – 523(K) / MMT – 723(K) Paradigm | unaffected |
| ||||||
| Medtronic | MMT – 551 / MMT – 751 MiniMed 530G | unaffected |
| ||||||
| Medtronic | MMT – 554 / MMT – 754 MiniMed Veo | unaffected |
| ||||||
| Medtronic | MMT- 508 - MiniMed pump | unaffected |
|
Configuration 1
Running on/with
- n/a
Configuration 2
Running on/with
- n/a
Configuration 3
Running on/with
- n/a
Configuration 4
- n/a
Running on/with
- n/a
Configuration 5
Running on/with
- n/a
Configuration 6
Running on/with
- n/a
Configuration 7
- n/a
Running on/with
- n/a
Configuration 8
Running on/with
- n/a
Configuration 9
Running on/with
- n/a
No data.
No Red Hat product state for this CVE.
No package ranges for this CVE.
Remediation
Vendor solution
The remote option is turned off in the pump by default.
Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic.
Medtronic has released additional patient focused information https://www.medtronic.com/security .
Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic.
Metrics
No CVSS v4.0 score for this CVE.
CVSS:3.1/AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:N/A:N
CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
AV:A/AC:M/Au:N/C:P/I:N/A:N
This CVE is not in the KEV list.
No CISA SSVC assessment for this CVE yet.
Estimated probability of exploitation in the wild in the next 30 days (FIRST EPSS). As of Oct 3, 2026.
Score over time
2021–2026- EPSS v1
- EPSS v5
- EPSS v2
- EPSS v3
Percentile over time
- EPSS v1
- EPSS v5
- EPSS v2
- EPSS v3
Table of values (10 key points)
Flat stretches are collapsed; showing up to 120 newest points.
| Date | Score | Percentile | Model |
|---|---|---|---|
| Oct 3, 2026 | 0.47% (0.00474) | 38.83th | v5 (v2026.06.15) |
| Sep 20, 2026 | 0.47% (0.00474) | 40.17th | v5 (v2026.06.15) |
| Jul 20, 2024 | 0.05% (0.00053) | 21.45th | v3 (v2023.03.01) |
| May 31, 2024 | 0.05% (0.00053) | 20.72th | v3 (v2023.03.01) |
| Jan 4, 2024 | 0.05% (0.00053) | 18.68th | v3 (v2023.03.01) |
| Mar 7, 2023 | 0.06% (0.00064) | 26.02th | v3 (v2023.03.01) |
| Mar 6, 2023 | 0.89% (0.00885) | 27.89th | v2 (v2022.01.01) |
| Feb 4, 2022 | 0.89% (0.00885) | 10.50th | v2 (v2022.01.01) |
| Feb 3, 2022 | 0.62% (0.00624) | 17.80th | v5 (v2026.06.15) |
| Apr 14, 2021 | 0.62% (0.00624) | 0.00th | v1 |
References (3)
- http://www.securityfocus.com/bid/105044 vdb-entryThird Party AdvisoryVDB Entry
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 Third Party AdvisoryUS Government Resource
| Link | Providers | Tags |
|---|---|---|
| http://www.securityfocus.com/bid/105044 | vdb-entryThird Party AdvisoryVDB Entry | |
| https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html | ||
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | Third Party AdvisoryUS Government Resource |
Change history (0)
No recorded changes yet.