Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints
Published Jul 2, 2018
7.1
HIGHCVSS 3.1
EPSS 1.32%
Description
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Affected products
-
- Version All versionsStatusaffectedConstraints-
- Version
-
- Version All versionsStatusaffectedConstraints-
- Version
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Medtronic | 2090 CareLink Programmer | unaffected |
| ||||||
| Medtronic | 29901 Encore Programmer | unaffected |
|
Running on/with
- n/a
No data.
No Red Hat product state for this CVE.
No package ranges for this CVE.
Remediation
Vendor solution
Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual:
* Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer. * Only connect the programmer to managed, secure networks. * Update the software on the programmer when Medtronic updates are available. * Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation. * Offline updates are available, contact your Medtronic representative for more information.
Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism.
Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800‑638‑1991.
After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information.
Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner.
Medtronic has released a security bulletin for the 2090 CareLink Programmer https://www.medtronic.com/security .
Metrics
No CVSS v4.0 score for this CVE.
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
AV:A/AC:L/Au:S/C:P/I:P/A:P
This CVE is not in the KEV list.
No CISA SSVC assessment for this CVE yet.
Estimated probability of exploitation in the wild in the next 30 days (FIRST EPSS). As of Oct 1, 2026.
Score over time
2021–2026- EPSS v1
- EPSS v2
- EPSS v3
- EPSS v5
Percentile over time
- EPSS v1
- EPSS v2
- EPSS v3
- EPSS v5
Table of values (12 key points)
Flat stretches are collapsed; showing up to 120 newest points.
| Date | Score | Percentile | Model |
|---|---|---|---|
| Oct 1, 2026 | 1.32% (0.01320) | 69.80th | v5 (v2026.06.15) |
| Jun 15, 2026 | 1.32% (0.01320) | 67.03th | v5 (v2026.06.15) |
| Jul 20, 2024 | 0.04% (0.00042) | 0.36th | v3 (v2023.03.01) |
| Mar 7, 2023 | 0.04% (0.00042) | 0.50th | v3 (v2023.03.01) |
| Mar 6, 2023 | 0.89% (0.00885) | 27.89th | v2 (v2022.01.01) |
| Sep 10, 2022 | 0.89% (0.00885) | 26.24th | v2 (v2022.01.01) |
| Apr 1, 2022 | 0.89% (0.00885) | 24.24th | v2 (v2022.01.01) |
| Feb 4, 2022 | 0.89% (0.00885) | 10.50th | v2 (v2022.01.01) |
| Feb 3, 2022 | 0.22% (0.00221) | 1.67th | v1 |
| Jan 6, 2022 | 0.22% (0.00221) | 1.63th | v1 |
| Sep 1, 2021 | 0.22% (0.00221) | 4.29th | v1 |
| Apr 14, 2021 | 0.22% (0.00221) | 0.00th | v1 |
References (3)
- https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01 Third Party AdvisoryUS Government Resource
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01
| Link | Providers | Tags |
|---|---|---|
| https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html | ||
| https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01 | Third Party AdvisoryUS Government Resource | |
| https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01 |
Change history (0)
No recorded changes yet.