Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible
Published Apr 30, 2018
6.5
MEDIUMCVSS 3.0
EPSS 0.78%
Description
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Affected products
-
- Version IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18StatusaffectedConstraints-
- Version
Default status is the baseline for the product, each version can override it (e.g. patched versions marked unaffected).
| Vendor | Product | Default status | Versions | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Philips | IntelliVue MX40 Patient Worn Monitor | n/a |
|
- < b.06.18
Running on/with
- n/a
No data.
No Red Hat product state for this CVE.
No package ranges for this CVE.
Remediation
No remediation recorded yet.
Metrics
No CVSS v4.0 score for this CVE.
No CVSS v3.1 score for this CVE.
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
AV:A/AC:L/Au:N/C:N/I:N/A:P
This CVE is not in the KEV list.
No CISA SSVC assessment for this CVE yet.
Estimated probability of exploitation in the wild in the next 30 days (FIRST EPSS). As of Oct 1, 2026.
Score over time
2021–2026- EPSS v1
- EPSS v5
- EPSS v2
- EPSS v3
Percentile over time
- EPSS v1
- EPSS v5
- EPSS v2
- EPSS v3
Table of values (11 key points)
Flat stretches are collapsed; showing up to 120 newest points.
| Date | Score | Percentile | Model |
|---|---|---|---|
| Oct 1, 2026 | 0.78% (0.00784) | 54.41th | v5 (v2026.06.15) |
| Sep 20, 2026 | 0.78% (0.00784) | 54.51th | v5 (v2026.06.15) |
| Jul 20, 2024 | 0.12% (0.00117) | 45.98th | v3 (v2023.03.01) |
| Oct 31, 2023 | 0.12% (0.00117) | 45.34th | v3 (v2023.03.01) |
| Sep 21, 2023 | 0.11% (0.00110) | 43.70th | v3 (v2023.03.01) |
| Sep 3, 2023 | 0.13% (0.00134) | 47.83th | v3 (v2023.03.01) |
| Mar 7, 2023 | 0.11% (0.00115) | 43.42th | v3 (v2023.03.01) |
| Mar 6, 2023 | 0.95% (0.00954) | 36.37th | v2 (v2022.01.01) |
| Feb 4, 2022 | 0.95% (0.00954) | 16.28th | v2 (v2022.01.01) |
| Feb 3, 2022 | 0.83% (0.00833) | 24.58th | v5 (v2026.06.15) |
| Apr 14, 2021 | 0.83% (0.00833) | 0.00th | v1 |
References (3)
- http://www.securityfocus.com/bid/100813 vdb-entryx_refsource_BIDThird Party AdvisoryVDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 x_refsource_MISCThird Party AdvisoryUS Government Resource
- https://www.usa.philips.com/healthcare/about/customer-support/product-security x_refsource_CONFIRMVendor Advisory
| Link | Providers | Tags |
|---|---|---|
| http://www.securityfocus.com/bid/100813 | vdb-entryx_refsource_BIDThird Party AdvisoryVDB Entry | |
| https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 | x_refsource_MISCThird Party AdvisoryUS Government Resource | |
| https://www.usa.philips.com/healthcare/about/customer-support/product-security | x_refsource_CONFIRMVendor Advisory |
Change history (0)
No recorded changes yet.