Back

HIGH

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6

Published Feb 15, 2018

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.

Affected products

Remediation

No remediation recorded yet.

Metrics

References (2)

Change history (0)

No recorded changes yet.

Sources
CVE.org / MITRE
Status PUBLISHED
Assigner icscert
Published Feb 15, 2018
Updated Aug 5, 2024
Reserved Aug 9, 2017
NVD
Status Modified
Modified Jun 17, 2026
Red Hat
Severity n/a
Public date n/a