Back

MEDIUM

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected

Published Mar 26, 2019

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Affected products

Remediation

No remediation recorded yet.

Metrics

References (1)

Change history (0)

No recorded changes yet.

Sources
CVE.org / MITRE
Status PUBLISHED
Assigner icscert
Published Mar 26, 2019
Updated Aug 6, 2024
Reserved Aug 22, 2014
NVD
Status Modified
Modified Jun 17, 2026
Red Hat
Severity n/a
Public date n/a